AED-A15 Automatic External Defibrillator
Function
AED is intended for defibrillation of patients with sudden cardiac arrest due to the development of ventricular fibrillation and ventricular tachycardia in medical and pre-medical care, as well as for public and home use.
Advantages
№ | Name | Amount, pcs. |
1. | Automatic external defibrillator А15, including: | 1 |
1.1 | Disposable electrodes | 1 |
1.2 | Power supply battery | 1 |
1.3 | Antiseptic alcohol sterile wipe | 1 |
1.4 | Shaving stick | 1 |
1.5 | Soft carrying bag | 1 |
1.6 | Memory card SD | 1 |
2. | Documents | 1 |
Specifications
Discharge | ||
Pulse shape | Bipolar (bi-phase) trapezoidal with compensation of the patient's impedance | |
Discharge energy | Adult mode: (170 – 200) J (±15 %) Child mode: (45 – 50) J (±15 %) | |
Mode | Semi-automatic | |
Electrocardiogram | ||
Leads | II (RA, LL) | |
Range of chest resistance values | (25 – 200) Ohm | |
Heart rate | (20 - 300) bpm | |
Heart Rate Error | 1 bpm | |
Recognition | Ventricular fibrillation (VF) | ≥0.2 mV |
Ventricular tachycardia (VT) | ≥160 bpm | |
Connection of electrodes | Sound confirmation | |
Frequency | (0.5 – 30) Hz | |
Control | ||
Basic | Lid opening button, Patient mode switch, "Discharge" button | |
Indicators | ||
Visual | Action indicator panel, status display (status of AED A15, batteries, temperatures), LED indicators (patient mode switch) | |
Sound | Speaker (voice prompts): Sound intensity at a distance of 1 m from the AED A15, not more than | |
Buzzer (beeps (CPR)): Sound intensity at a distance of 1 m from the AED A15, not more than | Acoustic zoomer frequency 2,300 ± 300 Hz |
Physical characteristics
Overall dimensions | Not more than (240 × 294 × 95) mm (W × H × D) |
Weight | No more than 2.65 kg with battery without electrodes |
Application environment
Operation / Standby mode | |
Temperature of | (0…43) °C |
Relative humidity | (5 – 95) % (without moisture condensation) |
Working height | (0 – 4,575) m |
Allowable drop height | 1 m |
Ingress protection | IP54 (IEC60529) |
Note: The standby mode informs about the periodic start of the self-test with the battery installed in the AND A15 | |
Storage (in transport packaging) | |
Temperature of | (20…60) °C |
Relative humidity | (5 – 95) % (without moisture condensation) |
Working height | (0 – 12,192) m |
Self-test | |
Temperature of | (0…43) °C |
Periodicity | Every 24 hours, once per week, once per month |
Results | The status display displays "O", or "X" |
Note: The self-test can only be started when the battery is installed |
Backup and data transfer
Standard | SD memory card, IR port |
Disposable electrodes
Adult / Pediatric electrodes | |
Shelf life | 2 years from the date of manufacture |
Electrodes | Disposable electrodes |
Location: | Adults: Anterolateral Pediatric: Anteroposterior |
Minimum number of discharges at maximum energy | 10 energy discharges 200 J |
Total electrode area | 120 см2 (±5 %) |
Minimum active (gel-coated) electrode area | 80 см2 |
Width of adhesive electrode ring | 11 mm (±5 %) |
Electrodes cable length | Not less than 2 m |
Environmental conditions | |
Temperature of | Operation: (0…43) °C Storage: (0…43) °C |
Relative humidity | (5 – 95) % (without moisture condensation) |
Battery Power
Type | LiMnO2, non-rechargeable, with a long life |
Voltage / Capacity | 15 V, 4,200 mAh |
Shelf life (in original packaging) | 2 years from the date of manufacture |
Standby time (when connected to AND A15) | 5 years from the date of manufacture |
Performance | At least 200 200 J energy discharges (excluding the period between CPR and defibrillation) or 10 hours of operation in monitoring mode at an ambient temperature of 20 °C |
Environmental conditions | |
Temperature of | Operation: (0…43) °C Storage: (0…43) °C |
Relative humidity | (5 – 95) % (without moisture condensation) |