SLE 5000


SLE 5000 artificial lung ventilation apparatus for newborns


Function

The SLE 5000 artificial lung ventilation apparatus is a high-end device, the result of the constant search for new solutions to achieve an effective combination of quality care for the patient and spent money on it. It is designed to replace and support the function of breathing in newborns, including those with ELBW and children of early age (from 300 g to 20 kg)


Advantages

  • Possibility of combining several ventilation modes (CMV+TTV, SIMV+TTV+PSV, HFO+CMV)
  • Data displaying on the built-in colour LCD display
  • Eliminating the possibility of spontaneous appearance of positive expiratory pressure (PEEP)
  • Non-invasive ventilation with INCA CPAP
  • Eliminates the pneumatic effect
  • Synchronizing modes and implementing the principles of "protective" ventilation
  • Constant pressure support at all frequencies
  • Exhaled gases filtration
  • Using gas mixtures of nitric oxide
  • Alarm system
  • Emergency shutdown of high pressure


Ventilation Modes

  • Convection mandatory ventilation of lungs (CMV)
  • Patient-initiated lung ventilation (PTV)
  • Synchronized intermittent mandatory ventilation (SIMV)
  • Support for breathing at constant positive pressure (CPAP), manual inhalation
  • Non-invasive ventilation with INCA CPAP
  • Reduction of volumetrauma and barotrauma, the possibility to switch to independent breathing (TTV)
  • Positive expiratory pressure (PEEP)
  • Auxiliary pressure support ventilation (PSV)
  • High-frequency ventilation (HFO)


Specifications

Traditional ventilation modes
CPAP / PTV / PSV
Inhalation time
0.1 - 3.0 sec.
CPAP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Specified volume
2 - 200 ml
FiO2
21% - 100%
CMV / SIMV
BPM (breathing rhythm)
1-150 inhalation/min.
Inhalation/expiration rate (I:E)
11.2:1 - 1:600
Inhalation time
0.1 - 3.0 sec.
PEEP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Specified volume
2 - 200 ml
FiO2
21% - 100%
Ventilation Modes: HFO High-frequency ventilation
Frequency range
3 - 20 Hz
Inhalation/expiration rate (I:E)
1:1
Delta Pressure Range
4 - 180 mbar
Mean pressure range
0 - 35 mbar
FiO2
21% - 100%
HFO+CMV
Breathing rate BPM
1 - 150 bpm
Inhalation time
0.1 - 3.0 sec.
Frequency range
3 - 20 Hz
Inhalation/expiration rate
1:1 / 1:2 / 1:3
Inhalation pressure
0 - 65 mbar
Delta Pressure Range
4 - 160 mbar
Mean pressure range
4 - 160 mbar
FiO2
21% - 100%
Monitoring Parameters of Flow and Volume Measurement
Flow sensor type - double heated anemometer (autoclavable or disposable)
10 mm
Flow (Accuracy±8%)
0.2 - 32 l / min
Expiratory tidal volume
0 - 999 ml
Expiratory minute volume
0 - 18 l
Dead spot
1 ml
Weight
10 g
Traditional Ventilation and Combined Modes:
Leakage in a patient
0 - 50% (Resolution - 5%, average for the last 5 breaths)
Total breathing rate
0 - 150 inhalation/min.
Dynamic compliance
0 - 100 ml/mbar (Resolution - 1 ml/mbar)
C20/C
resolution 0.1
Measurement time
2 msec
Resistance
0-1000 mbar.sec / l
Flow trigger sensitivity
0.2 - 10 l / min


Oxygen concentration

Range
21 - 100% (resolution 1%)


Pressure

Pressure measurement in real time
Resolution 1 mbar
Measurement time
2 msec
Peak pressure
0 - 175 mbar (Resolution - 1 mbar)
PEEP pressure
0 - 175 mbar (Resolution - 1 mbar)
Mean pressure
-175 - 175 mbar (Resolution - 1 mbar)


User-Defined Alarms

Peak pressure
Auto install, when the patient pressure control is adjustable or can be adjusted manually
Range
10 - 110 mbar
Resolution
0.5 mbar
Cycle Failure
Auto install, when the patient pressure control is adjustable or can be adjusted manually
Low pressure
Auto install, when the patient pressure control is adjustable or can be adjusted manually
Range
10 - 70 mbar
Traditional modes, HFO modes
up to 10 mbar below the high pressure threshold
Low Tidal Volume
Range
0 - 200 ml
Resolution
0.2 ml
Low Minute Volume
Range
from 0 to 0.02 liters below the High Minute Volume Threshold
Resolution
0.2 ml
High Minute Volume
Range
0.02 from 18 l
Resolution
0.1 l
Apnea Time
Set in CPAP or when the respiratory support rate is less than 20 BPM
Range
from 3 to 60 sec
Resolution
1 sec


Power

Voltage
100-250 V / 50-60 Hz
Power
115 VA
Battery
45-60 minutes
Battery charge
Full charge for 24 hours, 80% charge - for 8 hours


Outputs - RS-232C

Air and O2 connection

Pressure
3 - 5 bar
Fresh gas flow
8 l/min
Maximum gas flow
60 l/min


Application environment

Temperature of
10 - 40 ºC
Humidity
0-90% (without condensation)


Dimensions

Size, fan only
330mm (L) x 330mm (H) x 470mm (D)
Height on a short rack
114 cm
Height on a high rack
131 cm
Weight, fan without rack
21.8 kg

References