Reanimon

 


Newborn Respiratory Support Device


It provides noninvasive respiratory therapy for newborns, as well as short-term invasive and noninvasive ventilation through a T-shaped connector in resuscitation and intensive care units of maternity hospitals and perinatal centers.


Advantages

  • 8 modes of invasive and non-invasive ventilation
  • Suitable for both short-term emergency use and long-term ventilation
  • "T-contour" mode is a mode of manual invasive and non–invasive artificial ventilation of lungs during intensive care
  • High-precision regulation of the gas mixture supply even at low flows
  • Battery backup for at least 2 hours
  • Intelligent audio-visual alarm system
  • Monitoring of patient's parameters: pulse oximetry, heart rate
  • Apgar timer (for assessing the newborn's condition on the Apgar scale)
  • Metronome of cardiac and cardiopulmonary resuscitation (assistance with cardiac massage)
  • Quickly navigate through the interface menu using the rotary knob of the encoder
  • High-resolution touch screen display
  • High information content due to the large number of monitored parameters
  • Customizable trend display, 14-day data recording and storage


Technical specifications

General parameters
Patient categories
Newborns
The device is operable when connected to external oxygen and air sources at pressure
1.5 - 6 bar
The time when the device is ready for operation after switching on
no more than 30 s
Display type
TFT
Display diagonal
7”
Display resolution
1024х600
Sound and light alarms
available
Alarm sound level
at least 65 and no more than 85d BA
Protection class against electric shock
I
Protection from dust and moisture
IР22
Ventilation modes
Ventilation modes
T-contour, nCPAP, nCPAP Flow, biCPAP, biCPAP Flow, HFNC, LFNC, free О2
Patient monitoring
Measuring range Sp02
60 – 100 %

Permissible absolute error in measuring SpO2 values:

- in the range from 90% to 100% without patient movement

- in the range from 90 % to 100 % with patient movement

- in the range from 60 % to 89 %


± 2 %

± 3 %

± 3 %

The range of measurement and display of heart rate values
30 – 300 beats/min

Permissible absolute error in measuring heart rate values:

- without patient movement

- when the patient is moving


± 3 beats/min

± 5 beats/min
Overall dimensions and weight
Overall dimensions (Height x Width x Length)
300х250х230 mm
Weight of the device
6 kg
Power supply
Supply mains voltage
110/230 V
Supply mains frequency
50/60 Hz
Built-in battery capacity
4500 mAh
The charging time of the built-in battery from the mains
7 h
Built-in battery life
2 h
Power consumption
240 V·A
Components
The CPAP device
Basic
Oxygen hose for quick-release connection
Basic
Air hose for quick-release connection
Basic
Power cable
Basic
Breathing mixture humidifier TEVLAR
Additional
Respiratory circuit for intensive care T-contour
Additional
Breathing circuit for nCPAP
Additional
Holders-clamps for hoses and tubes-caps
Additional
Anatomical facial masks, CPAP masks, nasal cannulas
Additional
Neonatal pulse oximeter sensor
Additional
SpO2 adapter
Additional